The upkeep that keeps flue gas analyser readings trustworthy, and why each step matters to the safety decision.
A neglected analyser can make dangerous combustion look acceptable, and sensor drift is not predictably on the safe side. Work through these every time:
- Check calibration is in date. Electrochemical sensors age and lose sensitivity, so the analyser is calibrated against a known standard at the interval the manufacturer sets, commonly every 12 months, with a certificate as the record. If the certificate has lapsed, do not rely on the analyser for a safety decision: recalibrate it or use an in-date instrument.
- Replace sensors when due. Sensors have a limited service life and are replaced as the manufacturer advises, not automatically at every annual service.
- Inspect the sample path. Check the probe, hose, connections and seals for damage or leaks. A leaking sample path draws in room air, dilutes CO and CO2 and makes combustion look falsely clean.
- Empty the water trap and check the filter. Flue gas carries condensed water and soot. Water reaching the cells can damage them, and a full trap or clogged filter restricts the sample and gives slow or false readings. Empty the trap and maintain or replace the filter as the instrument instructions require. If liquid has got in, check for sensor damage: emptying the trap does not make the readings accurate again.
- Zero in clean outdoor air. Switch on away from the flue and vehicle exhaust and let the warm-up and purge countdown finish. The start-up sets the gas sensors' reference in clean air. Contaminated air at start-up throws off the gas readings and the figures calculated from them; the temperature and pressure readings are separate. Complete the instrument’s prescribed startup and checks before use.
- Wait for readings to settle after inserting the probe, so the sensors respond fully to the sample.
- Stop if readings are erratic. Investigate the instrument, setup and sample path, and check it works normally before you rely on a new result. Arrange service or repair where needed; an overdue instrument still requires its scheduled calibration. Do not use repeated tests to continue an unsafe exposure.
- Purge after use. Use its prescribed fresh-air shutdown purge to clear sampled gas from the sensing path. Drain collected liquid separately; a purge does not replace trap maintenance or repair water-damaged sensors.
Operating limits
Check which part of the equipment a limit describes. For the KANE458s:
- The analyser operates at ambient temperatures of 0–45 °C, with 15–90% non-condensing relative humidity.
- Its battery charger has separate 0–40 °C, indoor-use conditions.
- Its flue probe can measure up to 600 °C; that does not allow the analyser body to sit on a hot surface.
Use the conditions for the part being operated or charged. A probe rating is not the instrument's ambient rating.
Why the oxygen sensor matters
Many analysers calculate CO2, and so the CO/CO2 ratio, from the measured oxygen. A drifting oxygen sensor therefore puts an error into the ratio used to judge combustion.