Flue gas analyser readings glossary

Note 6 of 15 · free to read

Plain definitions of the figures on a flue gas analyser display, with the units and normal values you need to know.

Gases and units

  • FGA — flue gas analyser: the instrument used to check combustion.
  • Oxygen (O2) — the oxygen in the sample, shown as a percentage. Clean outdoor air reads about 20.9%. In flue gas, more oxygen means more excess air or dilution, and less oxygen means less. Where the instrument calculates CO2 from O2 and fuel selection, oxygen-sensor error also affects the derived CO2 figure and ratio. Other instruments measure CO2 directly.
  • Carbon dioxide (CO2) — produced when carbon burns fully, shown as a percentage. It can appear in a sample that also contains CO, so finding it is not a pass. Depending on the instrument, CO2 is measured directly or calculated from oxygen and the selected fuel; check which yours does.
  • Carbon monoxide (CO) — the toxic product of incomplete combustion, shown in parts per million because it is harmful at very low concentrations. Low CO can also come from a diluted or faulty sample, so validate the sample before judging combustion.
  • ppm — parts per million: 10,000 ppm = 1%, so 94 ppm is 0.0094%.
  • CO/CO2 ratio — CO divided by CO2 in matching units. A lower result means less CO relative to CO2; compare both the ratio and the measured CO with the appliance limits. A low ratio does not mean the installation is safe.

Air and dilution

  • Excess air — air supplied above the ideal amount, seen through the O2 and CO2 readings. For a given fuel, additional air generally raises O2 and lowers CO2. Reducing excess air is not an unlimited improvement: insufficient air or poor mixing can raise CO. Use the specified range and all required combustion readings.
  • Dilution — extra air getting into the sample through a leaking probe or hose or a wrong sampling point. O2 rises and CO2 and CO fall, so combustion looks better than it is.
  • Why the ratio resists dilution — both concentrations generally fall together, so their ratio changes less than raw CO. Background gas, weak samples and instrument uncertainty limit this approximation; it does not make a leaking sample valid.

Temperatures and efficiency

  • Flue gas temperature — the temperature of the products leaving the appliance. An unusually high flue-to-inlet-air temperature difference can indicate greater exhaust losses or poor heat transfer. Compare like-for-like firing and water conditions, and check the temperature sensors and probe position before diagnosing the cause.
  • Ambient or inlet temperature — the reference air temperature used with flue temperature in efficiency calculations. Use the instrument's specified sensor and position.
  • Net and gross efficiency — net (lower heating value) leaves out the latent heat of the water vapour made by combustion; gross (higher heating value) includes it. For the same useful heat and fuel quantity, the net basis gives a higher percentage. An analyser’s calculated combustion efficiency is not a direct measurement of the whole boiler’s useful or seasonal efficiency; check which measure and basis are being compared.

Instrument terms

  • Fresh-air zero — the start-up in clean air that sets the sensor baseline. It is needed before each use and does not replace periodic calibration.
  • Calibration — the scheduled check against a known standard that corrects for sensor drift, recorded on a certificate.