Interpret %LEL, account for sample-tube delay and air ingress, and recognise sensor faults, oxygen dependence and model-specific calibration factors.
Choose the measurement
Concentration describes how much gas is present in the sampled mixture. An instrument may display parts per million, percentage by volume or percentage of the lower explosive limit (%LEL). Check which quantity it measures and which gases it can detect.
An adjustable ticking leak seeker is useful for tracing a source, but its tick rate is not a calibrated %LEL measurement. To record that quantity, use a suitable concentration monitor with the correct target-gas setting, range and calibration. A pressure gauge or gas meter measures something else.
These principles are for professional portable instruments. They do not cover where to put household alarms, and they are not a confined-space entry procedure.
Get the intended sample to the sensor
A diffusion instrument relies on gas reaching its exposed sensor. An aspirated instrument uses a pump to bring a sample through a probe or tube. A diffusion sensor cannot pull a sample from the far end of an extension line.
Use the approved sampling arrangement. Inspect the tube, connections, filter and instrument before use. Kinks, blockages, water and damage can stop a true sample reaching the sensor.
A crack on the suction side of a pumped line lets in surrounding air. That dilutes the sample, so the detector sees less gas than there is at the sampling point. Repair the line and repeat the check as instructed. A steady number on the display does not mean the sample was taken correctly.
Allow time for the complete system
Gas takes time to travel through a sample tube. The sensor and electronics also take time to respond. A longer tube can therefore add delay even when the sensor itself is fast.
A sensor’s T90 figure is the time it takes to reach 90% of its final reading after a sudden change at its sample point. It does not include the time gas takes to travel through a separate extension tube. Follow the instrument’s instructions for the actual tube and set-up before you accept a reading at a new position.
Check whether the result is valid
A zero shown with a sensor-fault warning cannot be trusted: a failed detector can show zero while gas is present. Follow its fault procedure and the site’s safety arrangements, and get working monitoring back before you rely on any result.
Sensor type matters too. A catalytic pellistor uses the heat produced when gas reacts with oxygen at a heated bead. Too little oxygen can upset that reaction and make the gas reading wrong. That is a limit of the sensing method, not a way of telling whether the air is safe to breathe. Not every detector uses a pellistor, and their operating limits differ.
Match calibration to the gas
A detector can respond differently to two gases at the same concentration. If the target gas differs from its calibration gas, use the manufacturer’s approved setting or response correction for the fitted sensor and model. Never borrow a factor from another instrument just because both detect combustible gas.
Relative density, calorific value and meter-volume correction measure other things; none of them gives the detector’s response factor.
Understand the %LEL scale
The lower explosive limit is the reference point for %LEL. A reading of 100%LEL means that reference concentration has been reached. It does not mean the sample is 100% fuel gas, and it does not refer to the upper explosive limit or maximum flame speed.
Use the calibration basis specified for the instrument. For example, where 100%LEL methane corresponds to 4.4% by volume:
- 25%LEL is one quarter of that reference.
- One quarter of 4.4% is 1.10% methane by volume.
- 50%LEL on the same basis is 2.20% by volume.
The arithmetic explains the scale. It does not set an alarm level or make it safe to work in that atmosphere: follow the monitoring and emergency procedures.